Article Text
Abstract
Objective To assess feasibility of overnight closed-loop therapy in young children with type 1 diabetes and contrast closed loop using diluted versus standard insulin strength.
Research design and methods Eleven children (male 6; age range 3.75–6.96 years; glycated hemoglobin 60 (14) mmol/mol; body mass index SD score 1.0 (0.8); diabetes duration 2.2 (1.0) years, mean (SD); total daily dose 12.9 (10.6, 16.5) IU/day, median (IQR)) were studied at a clinical research facility on two occasions. In random order, participants received closed loop with diluted insulin aspart (CL_Dil; 20 IU/mL) or closed loop with standard aspart (CL_Std; 100 IU/mL) from 17:00 until 8:00 the following morning. Children consumed an evening meal at 17:00 (44 (12) gCHO) and an optional bedtime snack (6 (7) gCHO) identical on both occasions. Meal insulin boluses were calculated by standard pump bolus calculators. Basal rates on insulin pump were adjusted every 15 min as directed by a model-predictive-control algorithm informed by a real-time glucose sensor values.
Results Mean plasma glucose was 122 (24) mg/dL during CL_Dil vs 122 (23) mg/dL during CL_Std (p=0.993). The time spent in the target glucose range 70–145 mg/dL was 83 (70, 100)% vs 72 (54, 81)% (p=0.328). Time above 145 mg/dL was 13 (0, 27)% vs 19 (10, 45)% (p=0.477) and time spent below 70 mg/dL was 0.0 (0.0, 1.4)% vs 1.4 (0.0, 11.6)% (p=0.161). One asymptomatic hypoglycemia below 63 mg/dL occurred in one participant during CL_Dil versus six episodes in five participants during CL_Std (p=0.09). Glucose variability measured by CV of plasma glucose tended to be reduced during CL_Dil (20% (13, 31) vs 32% (24, 42), p=0.075).
Conclusions In this feasibility study, closed-loop therapy maintained good overnight glucose control with tendency towards reduced hypoglycemia and reduced glucose variability using diluted insulin.
Trial registration number clinicaltrials.gov Identifier: NCT01557634.
- Artificial Pancreas
- Type 1
- Child
This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
Statistics from Altmetric.com
Supplementary materials
Supplementary Data
This web only file has been produced by the BMJ Publishing Group from an electronic file supplied by the author(s) and has not been edited for content.
Files in this Data Supplement:
- Data supplement 1 - Online supplement