Table 2

Comparisons at various glucose concentrations of Eversense versus FSL versus FSLCstrip measurements as reference

Range
(mg/dL)
NFSLCstrip (1)Eversense (2)FSL
(3)
Exercise
≤706158±1059±1573±17**
71–180544125±30117±39**142±43**
>180303241±53222±59**276±76**
Total908159±72148±71**182±88**
Normal daily activity
≤704959±968±26*60±13
71–180504127±30122±33**128±39
>180261238±47223±51**240±57
Total814158±67151±64**160±73
  • 2 vs 1 and 3 vs 1.

  • *P<0.05; **P<0.001.

  • FSL, Free Style Libre; FSLCstrip, Free Style Libre Precision NeoPro strip.