Evaluation of turbidimetric inhibition immunoassay (TINIA) and HPLC methods for glycated haemoglobin determination

J Clin Lab Anal. 2012 Nov;26(6):481-5. doi: 10.1002/jcla.21550.

Abstract

Background: Various factors may affect the accuracy of hemoglobin (Hb) A1c measurements that are widely used to monitor glycemic control in diabetic patients. This study was aimed to compare the values of HbA1c obtained by two different methods, Roche Tina-quant second and thirdgeneration HbA1c assays based on the turbidimetric inhibition immunoassay (TINIA), and high-performance liquid chromatography (HPLC) cation-exchange method used by Arkray Adams HA-8160 analyzer.

Methods: Measurements of HbA1c were carried out in blood samples from 2,917 patients using above-mentioned methods. Linear regression was used for the correlation analysis and linear equations. Bland-Altman plots were performed from method comparison data using MedCalc statistical software.

Results: For the low control, the second generation Tina-quant assay had within-run and between-run CVs 0.8% and 0.9%; for the high control within-run and between-run CVs were 1% and 0.96%, respectively. HPLC method for the low control had within-run CV 1% and between-run CV 1.3%; for the high control within-run CV was 0.6% and between-run CV was 0.9%.

Conclusion: There was a good concordance between the results of TINIA and HPLC methods (y = 1.091x - 0.363; r(2) = 0.96).

Publication types

  • Comparative Study

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Glycated Hemoglobin / analysis*
  • Humans
  • Immunoassay / methods*
  • Linear Models
  • Nephelometry and Turbidimetry
  • Reproducibility of Results

Substances

  • Glycated Hemoglobin A