Accuracy of two continuous glucose monitoring systems: a head-to-head comparison under clinical research centre and daily life conditions

Diabetes Obes Metab. 2015 Apr;17(4):343-9. doi: 10.1111/dom.12378. Epub 2014 Sep 10.

Abstract

Aims: To assess the accuracy and reliability of the two most widely used continuous glucose monitoring (CGM) systems.

Methods: We studied the Dexcom®G4 Platinum (DG4P; Dexcom, San Diego, CA, USA) and Medtronic Paradigm Veo Enlite system (ENL; Medtronic, Northridge, CA, USA) CGM systems, in 24 patients with type 1 diabetes. The CGM systems were tested during 6-day home use and a nested 6-h clinical research centre (CRC) visit. During the CRC visit, frequent venous blood glucose samples were used as reference while patients received a meal with an increased insulin bolus to induce an aggravated postprandial glucose nadir. At home, patients performed at least six reference capillary blood measurements per day. A Wilcoxon signed-rank test was performed using all data points ≥15 min apart.

Results: The overall mean absolute relative difference (MARD) value [standard deviation (s.d.)] measured at the CRC was 13.6 (11.0)% for the DG4P and 16.6 (13.5)% for the ENL [p < 0.0002, confidence interval of difference (CI Δ) 1.7-4.3%, n = 530]. The overall MARD assessed at home was 12.2 (12.0)% for the DG4P and 19.9 (20.5)% for the ENL (p < 0.0001, CI Δ = 5.8-8.7%, n = 839). During the CRC visit, the MARD in the hypoglycaemic range [≤3.9 mmol/l (70 mg/dl)], was 17.6 (12.2)% for the DG4P and 24.6 (18.8)% for the ENL (p = 0.005, CI Δ 3.1-10.7%, n = 117). Both sensors showed higher MARD values during hypoglycaemia than during euglycaemia [3.9-10 mmol/l (70-180 mg/dl)]: for the DG4P 17.6 versus 13.0% and for the ENL 24.6 versus 14.2%.

Conclusions: During circumstances of intended use, including both a CRC and home phase, the ENL was noticeably less accurate than the DG4P sensor. Both sensors showed lower accuracy in the hypoglycaemic range. The DG4P was less affected by this negative effect of hypoglycaemia on sensor accuracy than was the ENL.

Trial registration: ClinicalTrials.gov NCT01751932.

Keywords: CSII; type 1 diabetes.

Publication types

  • Comparative Study
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Activities of Daily Living
  • Adult
  • Austria
  • Biomedical Research / instrumentation
  • Blood Glucose / analysis*
  • Diabetes Mellitus, Type 1 / blood*
  • Diabetes Mellitus, Type 1 / drug therapy
  • Female
  • France
  • Humans
  • Hyperglycemia / diagnosis*
  • Hyperglycemia / prevention & control
  • Hypoglycemia / chemically induced
  • Hypoglycemia / diagnosis*
  • Hypoglycemia / prevention & control
  • Hypoglycemic Agents / administration & dosage
  • Hypoglycemic Agents / therapeutic use
  • Injections, Subcutaneous
  • Insulin / administration & dosage
  • Insulin / therapeutic use
  • Insulin Infusion Systems
  • Italy
  • Male
  • Materials Testing
  • Middle Aged
  • Monitoring, Ambulatory / instrumentation*
  • Netherlands
  • Reproducibility of Results

Substances

  • Blood Glucose
  • Hypoglycemic Agents
  • Insulin

Associated data

  • ClinicalTrials.gov/NCT01751932